Culture Myograph System Model 202CM
7.2 Mounting Protocol for Small Arteries
This section shortly describes the basic technique of mounting small vessels in the
culture myograph:
One end of the vessel is care-fully mounted and secured (with two ne nylon
sutures) to one of the hollow glass micro-cannulas.
The lumen is ushed gently with PSS to remove any blood or debris.
The second end of the vessel is then mounted and secured to the second
micro-cannula.
The PSS in the myograph chamber is gradually warmed to 37
0
C over a periode
of 15 minutes.
Figure 7.6 Mounting of vessel in the
culture myograph
Figure 7.7 Vessel secured to the two glass
Seitenansicht 53