5.1.2 Removal and Mounting of Glass Cannulas and Chamber
Replacement of glass cannulas and demounting of the culture myograph chamber for
sterilisation are common routines in culture myograph experiments. The procedure is
illustrated in Fig. 5.3.
Removal of Left Glass Cannula:
Carefully remove the two O-rings (marked by the two arrows) to free the glass cannula
from the culture myograph unit. Then carefully unscrew the locknut and remove both
locknut and glass cannula.
Removal of right Glass Cannula:
Gently remove the little screw (marked by the arrow) to free the glass cannula from
the culture myograph unit. Then carefully unscrew the locknut and remove both
locknut and glass cannula.
Removal of Culture Myograph Chamber:
After removal of both glass cannulas, loosen the horizontal adjustment screw to free
and remove the chamber from the culture myograph chamber.
Mounting of all parts is performed the same way, but in opposite direction.
Fixation screw
O-rings
Lock-screw
Figure 5.3 How to free the glass cannulas from the culture myograph unit
43
User Manual vers. 1.4 Chapter 5