Culture Myograph System Model 202CM
Data logging is not automatic and must be activated by selecting (or creating) a le.
Files are chosen by clicking the “start data logging” button in the data control bar
or by selecting the “start logging data” menu item in the pop-up menu (click on )
on the data control bar. When data logging is active the “cross” indicator becomes
a “tick” indicator . Data logging can be stopped at any time by selecting the “stop
data logging” button in the data control bar or by selecting the “start/stop logging
data” menu item in the pop-up menu. Note that data will be stored in the data history
graph regardless of whether data is logged to disk.
3.3.4 Analysis
The analysis control bar is used to activate and determine the type of analysis to
be performed on the acquired images. The purpose of this analysis is to determine
the dimensions of the vessel in the myograph and to derive other parameters (e.g.
vessel cross section) based upon those dimensions. Although the video microscope
supplies the image for this analysis, some judgements must be made concerning
the analysis region and the nature of the analysis. VediView
uses a unique vessel
detection and analysis (VDA) methodology to recognise the presence of a vessel and
locate its edges. This method is largely automatic and effectively checks the correct
imaging of the vessel and is capable of estimating the position of lumen edges even
when the vessel is contracting (when conventional edge detection can not be used).
Since the analysis is implemented in real time on live images, the user can interact
with and recongure the analysis so that an optimum setting can be achieved for the
given vessel and microscope being used. VediView
provides this feature to the user
without them requiring any specialist knowledge of the image processing algorithms
being implemented. The pop-up menu on the control bar (select ) can be used to
access all functions of the analysis bar. The pop-up menu can also be revealed by a
right mouse button click when the Windows cursor is above the control bar. A more
detailed description of the functions available in the analysis control bar follows
under these sub-headings:
Setting the Analysis Rectangle
Calibration of Image Analysis
Starting Real Time Analysis
Conguration of Vessel Detection and Analysis (VDA)
Logging and Reviewing Results of Analysis
3.3.4.1 Setting the Analysis Rectangle
If the user requires a calibrated image analysis, the dimensions of the image must
be determined before commencing analysis. To perform this calibration the user
is referred to How To... Calibrate the Image Analysis. It is now assumed that the
calibration (if required) has already been performed. Before selecting a region for
analysis, the image should be displayed in the main window. Drag and click the
mouse to select a rectangle in the main window and then select the “Set analysis
rectangle” icon on the tool bar. The analysis rectangle may be re-selected by
simply drawing a second rectangle and re-selecting the “Set analysis rectangle” icon
. These processes may be repeated until the user is satised with the selected
region.
Figure 3.16
Analysis control bar