5.3 Pressure Regulator
The Pressure Regulator is connected to
the culture myograph system to generate a
pressure on the mounted vessel from the
perfusion buffer. The pressure is adjusted using
the knob and air release valve shown in Fig. 5.6.
5.4 Culture Myograph Maintenance
The Culture Myograph System – Model 202CM is a very delicate and sophisticated
piece of research equipment. In order to keep it working at its best, DMT recommend
that the following sections are read carefully and that the instructions are followed at
all times.
DMT strongly recommends that the myograph chamber and surroundings be cleaned
after each experiment.
After a “normal” experiment use the following procedure to clean the myograph
chamber and glass cannulas:
1. Fill up the myograph chamber to the edge with an 8% acetic acid solution and
allow it to stand for a few minutes to dissolve calcium deposits and other salt
build-up. Use a swab stick to mechanically clean all chamber surfaces.
2. Remove the acetic acid and wash the myograph chamber and glass cannulas
several times with double distilled water.
3. It is hard to remove any kind of hydrophobic reagent used by using step
1. and 2., try incubating the chamber and glass cannulas with 96% ethanol or
a weak detergent solution (e.g. Treepol).
4. To remove more resistant or toxic chemicals, incubate the myograph chamber
and glass cannulas with 1M HCl for up to 1 hour. In exceptional cases incubate
the chamber and supports with a up to 3M HNO
3
solution for about
15 minutes.
5. Wash the myograph chamber and glass cannulas several times with double
distilled water.
Important Notes:
To sterilize the culture myograph chamber and glass cannulas, use a standard
autoclave procedure.
• Be very careful using step c) and d) repeatedly as strong reagents can cause
extreme damage to the myograph unit.
• Be very careful not to damage the glass cannulas during the cleaning procedure.
Figure 5.6 Pressure Regulator
46
Culture Myograph System Model 202CM